![QuadTech Sentry Plus Series Скачать руководство пользователя страница 25](http://html2.mh-extra.com/html/quadtech/sentry-plus-series/sentry-plus-series_instruction-manual_3373286025.webp)
Introduction
Page 25 of 85
WARNING
MAKE SURE THE UNIT HAS BEEN DISCONNECTED FROM ITS AC POWER SOURCE
FOR AT LEAST FIVE MINUTES BEFORE PROCEEDING.
Procedure For Changing A Sentry Plus Series Fuse
Remove the fuse drawer, by pressing the black tab located at the center of the extended
fuse drawer, just below the 3-prong receptacle, and pull outward.
Once the fuse drawer has been removed from the instrument slide out the fuse from the holder
and replace. Make sure the new fuse is of the proper rating. Note that the fuse drawer can also
be used to store a spare fuse.
Install the fuse drawer back in the inlet module by pushing in until it locks securely in place.
Figure 1-5: Close-Up of Sentry Plus Series Rear Panel
1.4.4
Safety Inspection
Before operating the instrument inspect the power inlet module on the rear of the Sentry Plus to
ensure that the properly rated fuse is in place, otherwise damage to the unit is possible. Make
sure that the voltage selector switches are set in accordance with the power source in use. Refer
to paragraph 1.4.3 and Figure 1-5.
The Sentry Plus instrument is shipped with a standard U.S. power cord, QuadTech P/N 4200-
0300 (with Belden SPH-386 socket or equivalent, and a 3-wire plug conforming to IEC 320).
Make sure the instrument is only used with these cables (or other approved international cord
set) to ensure that the instrument is provided with connection to protective earth ground.
The surrounding environment should be free from excessive dust to prevent contamination of
electronic circuits. The surrounding environment should also be free from excessive vibration.
Do not expose the Sentry Plus instrument to direct sunlight, extreme temperature or humidity
variations, or corrosive chemicals.
100V/120V/220V/240V~
50/60Hz 300W MAX
VOLTAGE SELECTOR
90V - 110V ~
216V - 250V ~
198V - 242V ~
108V - 132V ~
3.15AT
250V
1.6AT
250V
FUSE Drawer