Faults
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
- 43 -
Faults
Calculation Errors
E400 C1
Calculation
Over
Range
E411 C2
The Calculation reading is greater than the specified upper limit, check channel
settings, calculation configuration, sensor condition and connections. If the
message persists please consult with your supplier.
E401
C1
Calculation Under Range
E411 C2
The Calculation reading is less than the specified lower limit, check channel
settings, calculation configuration, sensor condition and connections. If the
message persists please consult with your supplier.
E402
C1
Calculation Setup Checksum
E412 C2
The Setup for this Calculation has become corrupted. Check and correct the
calculation settings and turn the unit off and on again. If the message persists
please consult with your supplier.
E403
C1
Calculation Store A Checksum
E413 C2
The Store A Save for the channel associated with this calculation has become
corrupted. Check the calculation’s settings in the calculation menu and then save
the settings again in the Channel's Store A in the Save/Restore menu.
E404
C1
Calculation Store B Checksum
E414 C2
The Store B Save for the channel associated with this calculation has become
corrupted. Check the calculation’s settings in the calculation menu and then save
the settings again in the Channel's Store B in the Save/Restore menu.
E405
C1
Calculation SD Card Checksum
E415 C2
The SD Card store from which this Calculation was restored from has become
corrupted. Check the Calculation's settings in the Calculation menu and then save
the settings again to the SD card store.
Содержание MXD70 Series
Страница 1: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Страница 2: ......
Страница 6: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Страница 8: ...Specification 6 MXD70 Suspended Solids Turbidity Setup and Operating Guide Specification Blank ...
Страница 26: ...Channel Calibration 24 MXD70 Suspended Solids Turbidity Setup and Operating Guide Channel Calibration Blank ...
Страница 28: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Страница 49: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Страница 51: ......
Страница 52: ......