Faults
- 44 -
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
Faults
Modbus Errors
E420
UNIT
Modbus Setup Checksum
The Modbus setup has become corrupted. Check and correct the Modbus settings
and turn the unit off and on again. If the message persists please consult with your
supplier.
E421
UNIT
Modbus Store A Checksum
The Store A Save for the Modbus Configuration has become corrupted. Check the
Modbus settings in the Modbus menu and then save the settings again into Store A
in the Save/Restore menu.
E422
UNIT
Modbus Store B Checksum
The Store B Save for the Modbus Configuration has become corrupted. Check the
Modbus settings in the Modbus menu and then save the settings again into Store B
in the Save/Restore menu.
E423
UNIT
Modbus SD Card Checksum
The SD Card store from which the Modbus configuration was restored from has
become corrupted. Check the Modbus settings in the Modbus menu and then save
the settings again to the SD card store.
Содержание MXD70 Series
Страница 1: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Страница 2: ......
Страница 6: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Страница 8: ...Specification 6 MXD70 Suspended Solids Turbidity Setup and Operating Guide Specification Blank ...
Страница 26: ...Channel Calibration 24 MXD70 Suspended Solids Turbidity Setup and Operating Guide Channel Calibration Blank ...
Страница 28: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Страница 49: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Страница 51: ......
Страница 52: ......