Faults
- 46 -
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
Faults
The sensor Reading is Incorrect
Ensure the sensor is mounted properly, that there is at least 25mm (1”) of clearance around the head
of the probe.
Establish that the sensor is specified to work within the range that is being monitored.
Current Output Is Incorrect or Noisy
Check that the unit is “On-Line” (see page 10)
Check that the maximum load for the current loop has not been exceeded. (750
).
Check that the terminals have been wired correctly.
Check that the cable screen is attached to Earth at one end and that the cable does not pass too
close to a power cable.
Check that he current output has been configured properly.
Relays Appear to Malfunction
Check that the unit is “On-Line” (see page 10)
Check that the set point configuration is correct (see Setpoints, Current Outputs and Digital Input
Configuration Guide)
If the relays are vibrating or “chattering” as they pass the set point, check the hysteresis setting and
increase if necessary.
Ensure that the relays are connected properly (see MXD73 Installation guide or MXD75 Installation
guide) and that the voltage/current levels are not exceeding
5A @ 30V DC or 5A @ 250V AC.
Check that the instrument input cables are not picking up excessive noise.
Guarantee and Service
Products manufactured by Quadbeam Technologies Ltd are guaranteed against faulty workmanship
and materials for a period of three years from the date of despatch, except for finished goods not of
Quadbeam Technologies manufacture, which are subject to a separate agreement.
Goods for attention under guarantee (unless otherwise agreed) must be returned to the factory
carriage paid and, if accepted for free repair, will be returned to the customer’s address free of charge.
Arrangements can also be made for repair on site, in which case a charge may be made for the
engineer’s time and expenses.
If any services other than those covered by the guarantee are required, please contact Quadbeam
Technologies direct.
N.B. Overseas users should contact their Quadbeam Technologies nominated representative. Special
arrangements will be made in individual cases for goods returned from overseas.
Содержание MXD70 Series
Страница 1: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Страница 2: ......
Страница 6: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Страница 8: ...Specification 6 MXD70 Suspended Solids Turbidity Setup and Operating Guide Specification Blank ...
Страница 26: ...Channel Calibration 24 MXD70 Suspended Solids Turbidity Setup and Operating Guide Channel Calibration Blank ...
Страница 28: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Страница 49: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Страница 51: ......
Страница 52: ......