Specification
MXD70 Suspended Solids / Turbidity
Setup and Operating Guide
- 5 -
Specification
Suspended Solids Input Card Specification
Supported Sensor Types
Quadbeam S Series.
Sensor Input
Proportional probe signal from 0 to 16000
Linearization
The incoming probe signal can be converted to standard
engineering units using one of two user definable
linearization curves consisting of up to 10 points.
Sensor Cable Length
Up to 100 meters
Display Units
User selectable from,
%, NTU, FTU, mg/l, g/l, ppm, ppt, EBC, OD
In ranges of 0-10.00, 0-100.0, 0 – 9.999, 99.99, 999.9 and
9999 (ranges available vary depending on which units have
been selected).
Repeatability
±10 Probe input signal.
Repeatability
± 0.1% of range.
Calibration Timer
Inbuilt calibration count down timer which will trigger an
alarm when calibration interval has expired.
Sensor Input filter
Adjustable filter that averages the sensor input over a user
selectable time (1 – 32 Seconds).
Turbidity Input Card Specification
Supported Sensor Types
Quadbeam T Series.
Sensor Input
Proportional probe signal from 0 to 32000
Linearization
The incoming probe signal can be converted to standard
engineering units using one of two user definable
linearization curves consisting of up to 10 points.
Sensor Cable Length
Up to 100 meters
Display Units
User selectable from,
%, NTU, FTU, mg/l, g/l, ppm, ppt, EBC, OD
In ranges of 0-10.00, 0-100.0, 0 – 9.999, 99.99, 999.9 and
9999 (ranges available vary depending on which units have
been selected).
Repeatability
±10 Probe input signal.
Repeatability
± 0.1% of range.
Calibration Timer
Inbuilt calibration count down timer which will trigger an
alarm when calibration interval has expired.
Sensor Input filter
Adjustable filter that averages the sensor input over a user
selectable time (1 – 32 Seconds).
Содержание MXD70 Series
Страница 1: ...MXD70 SERIES Multi parameter Monitor Suspended Solids Turbidity Setup and Operating Guide ...
Страница 2: ......
Страница 6: ...Contents 4 MXD70 Suspended Solids Turbidity Setup and Operating Guide Blank ...
Страница 8: ...Specification 6 MXD70 Suspended Solids Turbidity Setup and Operating Guide Specification Blank ...
Страница 26: ...Channel Calibration 24 MXD70 Suspended Solids Turbidity Setup and Operating Guide Channel Calibration Blank ...
Страница 28: ...Appendix 26 MXD70 Suspended Solids Turbidity Setup and Operating Guide Appendix Blank ...
Страница 49: ...Notes MXD70 Suspended Solids Turbidity Setup and Operating Guide 47 ...
Страница 51: ......
Страница 52: ......