USING THE ULTRASONIC PROCESSOR
Press the
|
key on the front panel
.
The screen will display the power rating and the
frequency of the Ultrasonic Processor, and the following control parameters.
AMPLITUDE:
Amplitude is the intensity of sonication and the only parameter that must
be set in order for the Ultrasonic Processor to be operational. The other control
parameters
–
Time and Pulse, do not have to be set for continuous operation. AMPL
Displays the percentage of amplitude that was previously selected. Press the
AMPL
key
and then use the numeric keys to enter the desired value. Then depress the
ENTER/REVIEW
key.
For example, to enter a value of 40% amplitude, first depress the
4
key, then the
0
key and then depress the
ENTER/REVIEW
key.
The screen will display:
The Ultrasonic Processor is now ready for continuous operation. To energize the
equipment, press the
START
key or the footswitch. To de-energize the equipment,
press the
STOP
key or release the footswitch. If the Time or Pulse* functions must be
used, refer to the appropriate paragraphs(s) below.
To increase or decrease the amplitude in small increments when the equipment is on,
depress the
AMPL
key to display the percentage of amplitude that was previously
selected, then depress the
or
key, as required.
CAUTION
Do not operate the power supply unless it is
connected to the converter.
NOTE
Any combination of functions can be
selected in any order. To clear an
erroneous entry press the CLEAR key.
TIME __:__:__
TEMP__ __°C
PULSE _ _ _ _
AMPL__ __%
TIME __:__:__
TEMP__ __°C
PULSE _ _ _ _
AMPL 40%
20