24
DIRECT PROBE SONICATION
Proper setup is important to optimizing the sonication amplitude
settings and processing times. The probe should be securely
held using the Qsonica stand and clamp. This clamp can be
secured to either the booster ring or bottom of the converter.
Probe depth must be adjusted to ensure adequate mixing and
sample circulation. Larger volumes and viscous samples will
process more quickly and effectively with the addition of a mixer
or stir bar.
Probe Depth Example:
Probe Options
:
Part #
Tip Diameter
Processing
Volume
Range*
Amplitude
4777
1.5" (38mm)
5 - 20L
100
µ
m
4697
1" (25mm)
1 - 5L
100
µ
m
4776
0.5" (12mm)
50 - 400ml
100
µ
m
*Note
: Processing volumes are application specific. There are
many variables (viscosity, concentration, etc.) that can affect the
min/max processing volumes and processing times.
In addition, the amplitude setting needed for an application must
be determined empirically.
Contact us for help with selecting the amplitude and time settings
to begin your optimization trial.
Note:
the booster cannot
be ed i h
dia
meter tip probe #4776.
Probe depth is volume
dependent.
Do not insert the
#4777 probe deeper
than indicated by the
arrow. Contact us for
guidance.
16