2 . 1 T R A N S P O R T A N D A S S E M B LY
The image above shows supply and fuse on the device‘s upper quoin.
2 . S TA R T O F O P E R AT I O N
12
power inlet
fuse carrier
applicator B
applicator A
QRS
®
-101
and
QRS
®
-101 P
are mobile, mains-powered appliances that shall not be moved during intended
operation. Any location on a level surface is suitable. The devices should not be placed in front of a heater
or radiator (leave approx. 1 m clearance). There are no further requirements concerning wall distance or
ventilation.
The device meets the requirements DIN/VDE 0750, EN 60601-1, EN 60601-1-2 and belongs to the protection
class
I
. It is part of class IIa within the scope of the Medical Device Directive.
!
ATTENTION
The unit is not designed to be operated in places with the inherent risk of explosions. If it is
used in dangerous areas of anaesthesia departments, the possibility of an explosion cannot
be excluded.
!
ATTENTION
In order to prevent the risk of electric shock, the unit must be connected only to a supply
network with protective conductor.
If the patient and/or an applicator(-cable) is directly exposed to a radiator of a medical device for high
frequency heat therapy, the damage of the device or danger to the patient cannot be excluded. As a rule, a
distance of 3 m is suffi cient.
2 . 2 P O W E R S U P P LY
QRS
®
-101
and
QRS
®
-101 P
are intended to be connected to mains voltages of 115 V or 230 V and a mains
frequency of 50 or 60 Hz. Within this range there is no need for further change or alteration on the devices.
Connect the device’s power jack to a grounded socket by means of the attached power cord.