GB

Pressing the

A/B

OK

-key confi rms the current selection, leaves this menu and starts the device and its standard 

functionality.

!

NOTE

You may leave the menu to set the language at any time by pressing the  der 

Start 

Stop 

 -key. 

The currently made (incorrect) setting will then not be effective.

3 . 1 . 6   R E S T O R E   B A S I C   S E T T I N G S

Through a certain hot key you may restore the factory-made basic device settings.

You may restore these basic settings by

•  turning the device off,

•  pressing both

I/O

and

On/Off

-keys 

simultaneously, keeping them pressed, while the device 

  stays turned off

 and now 

•  turning the device on via the

On/Off

-key.

QRS

®

-101

 and

 QRS

®

-101 P 

will now reset to the following basic settings:

•  display light 

 

OFF

•  speaker 

 

ON

•  language 

 

DE (German)

Basic settings only for 

QRS

®

-101

•  therapy program 

 

relax

•  therapy intensity 

 

medium

•  therapy duration 

 

8 min

•  timer 1 

 

OFF and 06:00 o‘clock, „vital“, „medium“, 8 min.

•  timer 2 

 

OFF and 22:00 o‘clock, „relax“, „medium“, 8 min.

3 . 2   H O M E - T H E R A P Y   ( O N LY   Q R S

®

- 1 0 1 )

The  home  therapy  describes  a  mode  of  operation  where  a 

QRS

®

-101

 therapy is performed via simple key 

programming.

The operation is quite simple. For each therapy 3 parameters have to be defi ned:

•  the therapy program (relax, basis or vital),
•  the therapy intensity (sensitive, medium or intensive) and
•  the therapy duration (8, 16 or 24 minutes).

18

Содержание QRS-101

Страница 1: ...rs or omissions All rights reserved No part of this manual may be reproduced in any form or by any means electronic mechanical or otherwise without the prior written permission of the 456 Magnovit International Produktions und Handels AG 456 QWHUQDWLRQDO Sales 456 0DJQRYLW QGXVWULHVWUDVVH 5XJJHOO ZZZ TUV PDJQRYLW FRP RIILFH TUV PDJQRYLW FRP Manufacturer gbo Medizintechnik AG Kleiststraße 6 D 64668...

Страница 2: ...S 101 and QRS 101 P are electronic devices For their disposal the according regulations for electronic devices have to be observed Incidentals have to be disposed with residual waste On request the manufacturer will provide you with further technical descriptions for all repairable parts of the device such as circuit diagrams spare parts lists and adjustment instructions as far as these are necess...

Страница 3: ... Electrostatic discharge ESD IEC61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative h umidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lines Mains power qualit...

Страница 4: ...V m d 1 23P for 80 MHz to 800 MHz d 2 33P for 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts according to the transmitter manufacturer and d is the recom mended separation distance in meters m Interference may occur in the vicinity of equipment marked with the following symbol Recommended separation distances to portable and mobile RF communication equipm...

Страница 5: ...rapy Duration 19 3 3 Chip Card Mode 20 3 4 Timer Mode 21 4 APPLICATORS 23 4 1 The Coil Pillow K1 24 4 1 1 Dosage Guidelines for the Pillow Applicator 25 4 2 The Coil Mat M1 25 4 2 1 Application of Coil Mat M1 25 4 3 The Magnetic Field Bar MFS1 Pen Applicator 26 5 THE QRS MAGNETIC FIELD THERAPY 27 5 1 Three Reliable Programs Basis Vital and Relax Programs 28 6 TROUBLESHOOTING 29 6 1 Error Messages ...

Страница 6: ...ssories Chapter 5 explains how to carry out the therapies Chapter 6 lists possible failures their indication as well as their possible causes Chapter 7 states indications of safety related checks according to the MDD as well as routine maintenance Chapter 8 lists possible contraindications of QRS Therapy in addition to advices for avoiding hazards Chapter 9 states all of the device s relevant tech...

Страница 7: ...es to be treated as indicated Since the individual electromagnetic hypersensitivity varies depending on the patient s health and his blood s acidity the magnetic field intensity may be altered to achieve the ideal effect The magnetic field intensity must be explicitly chosen via operation QRS 101 and QRS 101 P work with field strengths up to 40 micro Tesla and therefore well below the WHO limit or...

Страница 8: ...QRS 101 QRS 101 P 1 2 VIEW OF QRS 101 A N D Q R S 101 P 10 GB ...

Страница 9: ... 2 selected therapy program e g the 2nd program on a chip card I O I O Start Stop 2 8 min 16 min 24 min 1 On Off 101 A B OK sensitive medium intensive relax basis vital timer 1 key key Í program key display backlight On Off Start Stop Taste On Off switch key Î applicator selection and confirmation key speaker On Off intensity and fine tuning therapy duration and fine tuning I O I O Start Stop 16 m...

Страница 10: ...The unit is not designed to be operated in places with the inherent risk of explosions If it is used in dangerous areas of anaesthesia departments the possibility of an explosion cannot be excluded ATTENTION In order to prevent the risk of electric shock the unit must be connected only to a supply network with protective conductor If the patient and or an applicator cable is directly exposed to a ...

Страница 11: ...t subject to aging or wear Thus a defect mains fuse always indicates an internal device error If the device does not show any function after being switched on display stays dark first make sure that the socket and mains cord conduct line voltage Should this be futile please check the mains fuse and replace it by a new fuse with the same values if it is defect The mains fuse is located in the fuse ...

Страница 12: ...apter 6 Troubleshooting If everything works properly The display will show If required please adapt application specific settings see chapter 3 1 The device is now ready for operation it displays the last operating state and waits for your input SYSTEMCHECK SYSTEM OK NOTE QRS 101 and QRS 101 P come with a time out turning the device off 8 minutes after not having been used 8 minutes after the end ...

Страница 13: ...al identification A or B but the descriptive logical identification mat pillow or pen You may for example see the following display This display will be shown for 1 5 seconds before returning to the display of the operating mode You may switch to the respective other applicator by pressing the A B OK key a second time while the above stated display is still visible PILLOW NOTE You may only switch ...

Страница 14: ...gnal status may be checked by simply pressing the I O key Depending on its status one of the following displays will appear These are always visible for 1 5 seconds before returning to the previous status If the I O key is being pressed while the display shown above is visible the end of therapy signal s status will be changed i e the switched on signal will be turned off and vice versa TH END BEE...

Страница 15: ... The currently made incorrect settings will then not be effective Only the German format for time and date are available exclusively 3 1 5 SETTING THE NATIONAL LANGUAGE You may operate QRS 101 and QRS 101 P in different languages To set the standard language the device s user interface offers you convenient menu guidance You may reach the corresponding menu by turning the device off pressing the A...

Страница 16: ...e stays turned off and now turning the device on via the On Off key QRS 101 and QRS 101 P will now reset to the following basic settings display light OFF speaker ON language DE German Basic settings only for QRS 101 therapy program relax therapy intensity medium therapy duration 8 min timer 1 OFF and 06 00 o clock vital medium 8 min timer 2 OFF and 22 00 o clock relax medium 8 min 3 2 HOME THERAP...

Страница 17: ...nto the card reader Please remove the chip card first The therapy intensity intensive is not available for the therapy program relax The intensity medium will always be chosen whenever the intensity intensive is being selected 3 2 1 FINE TUNING OF INTENSITY AND THERAPY DURATION When necessary the intensity and duration of the home therapy can be fine tuned The therapy duration can be adjusted star...

Страница 18: ...those with the keys 1 and 2 Following the therapy number the plain text therapy description will be displayed for example This plain text has been chosen by your physician or therapist and serves solely as mnemonical description to identify the therapy If this text is longer than 16 characters they will be displayed as a sort of ticker as running text Following the therapy description the respecti...

Страница 19: ...on is taken out of the power supply and reconnected later the device will switch on independently display the text AUTOMODE and switch off again The same display appears if a chip card is inserted into a switched off control unit 3 4 TIMER MODE ONLY QR S 1 0 1 QRS 101 contain 2 independent timers These programmable timing circuits serve to automatically start a therapy Here you are free to choose ...

Страница 20: ...change the time by increasing the digit positioned below the cursor by pressing the 2 key or to lessen it via the 1 key The A B OK key confirms the current value and goes to the next position After confirming the last time digit you automatically reach the menu to set the therapy parameters Alterna tely you can now see the settings programmed for this timer regarding therapy program therapy intens...

Страница 21: ... To vitalize and stabilize good health in healthy people we recommend 2 treatments per day about 8 hours apart The QRS system has several applicators at its disposal These guarantee the most effective therapy results even at the large variety of different indications One of the applicator specific characteristics is the corresponding magnetic flux density especially regarding the setting of a defi...

Страница 22: ... the body ATTENTION The magnetic field effectively spreads up to 0 3 meters sideways and up to 1 2 meters vertically which means that people in the immediate proximity during treatment will receive a gentle stimulation People not requiring treatment should remain outside the above mentioned range 24 ...

Страница 23: ...tes a magnetic field of different strength depending on the control unit settings For the Sensitive setting the magnetic field strength lies at about 0 3 µT for level intensive the magnetic field strength increases up to 30 µT The magnetic field is distributed over the whole coil mat uniformly Its impact ranges 0 3 meters beyond the mat and approx 1 2 meters above and below the mat The body is emb...

Страница 24: ... elbow for specific pain e g tennis elbow or for indications regarding the head eyes ears jaws and paranasal sinuses The selection of intensity and duration ensues the same as for the coil pillow Magnetic Acupuncture Especially well suited for children and adults who are afraid of needles Also suitable for acupuncture massage Toning one minute for each acupuncture point level intensive basic progr...

Страница 25: ...formation regarding the magnetic field strength settings is given in the preceding chapter Treatment times Possible treatments per day 3 times morning midday and evening After 6 PM the setting of the Basis program should not be higher than setting medium Use the Relax program to avoid negative influence on sleep Therapy duration per day Up to 2 hours is safe according to a study by Prof Dr Dr A Va...

Страница 26: ...n be treated especially organisms with a blood or lymph circulation The device is recommended for treating humans in the medical and sporting fields The Relax program Fields of Application Relaxation immunostimulation promotion of sleep To achieve these effects special frequencies for the circulatory system were reinforced in the Relax program compared to the Basis program It can be used with the ...

Страница 27: ...ected in a chip card therapy Please connect an applicator respectively the one prescribed on the chip card The chip card may not contain a therapy credit may be inserted the wrong way or may be mechanically damaged Please check if the chip card has been inserted correctly see label otherwise please contact your physician therapist service partner or distributor The applicator does not conduct curr...

Страница 28: ...rapy Please connect an applicator respectively the one prescribed on the chip card E102 The applicator does not conduct current The applicator or its connection cable may be defect usually caused by rough or inappropriate handling Please check if the applicator has been connected and tightened correctly If available test another applicator on the device s corresponding connection In case of visibl...

Страница 29: ...ontact your service partner A permanent alarm signal resounds after connecting the device to the mains power supply The control unit s program memory has the wrong content Please contact your service partner The menu to set the time appears every time the device is turned on The control unit s clock is powered by a battery its regular life span averages at more than 5 years The battery is empty an...

Страница 30: ... Medical Devices has to be especially observed NOTES Safety controls have to be made on a basis of the Ordinance on Operators of Medical Devices an ordinance connected with Medical Device Directive 93 42 EEC Please let us point out clearly that the Medical Device Directive is not valid outside the EC the Ordinance on Operators of Medical Devices is not valid for medical devices which serve neither...

Страница 31: ... the socket The device is suited for wiping disinfection It has to be observed that no liquids enter the device In no case shall the plug or socket get wet For cleaning or disinfection the device may not be drizzled Control Unit Applicators Do not immerse in water Use normal household cleaners for synthetic materials Only wipe with a damp cloth Be sure to keep sharp pointed objects away from mat a...

Страница 32: ...evice If the patient and or applicator or its connecting lead be located in the immediate sphere of a high frequency short wave or microwave therapy device the possible damage of device or patient may not be excluded Please keep a distance of at least 3 m The device is not intended to be used in areas with the risk of an explosion If it is used in anesthesia rooms in an endangered area a possible ...

Страница 33: ...h may also occur at the beginning of treatment please lower the intensity and or reduce the application to once a day In chronic illnesses initial aggravations may occur as they are known from homeopathic medicines The therapy should not be ended though Biological initial aggravations namely indicate the beginning reaction and or loosening of blockades within the body Possible second deterioration...

Страница 34: ...tection against ingress of water IPX0 Dimensions 92 mm x 190 mm x 215 mm h x w x d Weight 1 2 kg Display DOT matrix 16 characters Environmental conditions Operation of the device Temperature range 10 C up to 40 C Relative humidity 30 up to 75 Transport and storage Temperature range 5 C up to 50 C Relative humidity 90 none condensing Magnovit International Produktions und Handels AG reserves the ri...

Страница 35: ...t applicator M1 033 5 0010 Pillow applicator K1 033 5 0011 Magnetic field pen MFS1 033 5 0013 Carrying bag 033 8 0005 User manual 033 7 0013 Applicator glasses 033 5 0020 Applicator ear phone 033 5 0021 Power cable Europe 033 4 1001 Power cable Denmark 033 4 1002 Power cable Switzerland 033 4 1003 Power cable Great Britain 033 4 1004 Power cable Italy 033 4 1005 Power cable Australia 033 4 1006 Po...

Страница 36: ...n Transport 9 L Language 17 Local Application 26 M Magnetic Acupuncture 26 Magnetic Field Bar 26 Magnetic Flux Density 23 Mains Frequency 12 Mains Voltage 12 Mattress 25 33 Medical Device Directive 32 Medication 35 N Notified body 35 O Operating Staff Security 34 Ordinance on Operators of Medical Devices 32 P Pace Makers 34 Pain memory 27 Pregnancy 34 Protection degree 35 R Radiator 12 Relax progr...

Страница 37: ...GB NOTES 39 ...

Страница 38: ...GB 40 NOTES ...

Страница 39: ...GB NOTES 41 ...

Страница 40: ...served 456 QWHUQDWLRQDO 9HUWULHE HOWZHLW 456 0DJQRYLW QGXVWULHVWUDVVH 5XJJHOO RIILFH TUV PDJQRYLW FRP ZZZ TUV PDJQRYLW FRP Produktion Deutschland gbo Medizintechnik A OHLVWUDVVH D 64668 Rimbach QRS Production Germany QRS Produktion Deutschland ...

Отзывы: