EMC Tables
Guidance and manufacturer’s declaration –
electromagnetic emissions
The QRS
Opti
is intended for use in the electromagnetic environment specified below.
The customer or the user of the QRS
Opti
should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
The QRS
Opti
uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 N/A
The QRS
Opt
i uses batteries only and is not
connected to mains.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
N/A
Guidance and manufacturer’s declaration –
electromagnetic immunity
The QRS
Opti
is intended for use in the electromagnetic environment specified below.
The customer or the user of the QRS
Opti
should assure that it is used in such an
environment.
Emissions test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst IEC
61000-4-4
N/A
N/A
Surge IEC 61000-
4-5
N/A
N/A
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
N/A
N/A
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
N/A
N/A
Power frequency magnetic fields
should be at levels characteristic
of a typical commercial or hospital
environment.
11
10