Setting Up the System

Overview of the ABPM System

The basic components of the 

Opti

 ABPM system are:

• 

Opti

 ABP monitor with 4’ patient hose assembly

• 

OptiFit

 

cuffs (Adult and Adult Plus)

•  Pouch
•  Belt
• 

Opti

 User’s Guide

• 

Opti-Insight

 User’s Guide

• 

Opti-Insight

 CD

•  USB cable

See the Product and Accessories page for contents.

Hardware Requirements

•  Pentium based or equivalent PC-compatible computer with CD drive
•  SVGA or compatible display adapter and monitor (1024 x 768 resolution 

recommended)

•  One available USB or serial port
•  32 MB RAM and minimum 20 MB HDD recommended

Software Requirements

Microsoft Windows

®

 version 98 or later, excluding Windows ME

Microsoft Internet Explorer

®

  3 or later with Microsoft HTML Help. 

Powering the 

Opti

 for use

Install 2 AA batteries in the bay located at the back of the monitor. The label in the 
bay shows the orientation in which the batteries should be placed. When batteries are 
properly loaded, the monitor’s display will show the following:

1.  Incrementing dashes for two seconds
2.  Software and safety version of the monitor
3.  Battery voltage for two seconds
4.  Three quick beeps
5.  The number of BP readings in memory with flashing printer for three seconds
6.  One long beep
7.  Time flashing for twenty seconds

The monitor is now ready to be used.

Opti

 at a Glance

LCD & Key Diagram

START/STOP Button Functions Explained

Start/Stop 
Button

•  TO POWER ON: When the monitor is off, press the Start/Stop 

button.

•  TO POWER OFF: When the monitor is on but not taking a 

reading, press and hold the Start/Stop button until you hear five 
quick beeps, then release.

•  TO ABORT A MEASUREMENT: When the monitor is taking a 

reading and the cuff pressure is displayed, press the Start/Stop 
button.

•  TO START THE PROGRAMMED ABP STUDY: When the time is 

flashing, press the Start/Stop button.

•  TO START A SINGLE BP READING: When the time is displayed, 

press the Start/Stop button.

Time

Indicates current time; when flashing, the monitor will turn off in the 

next 20 seconds unless an ABP study is started.

Pressure

Indicates the pressure of the cuff in mmHg during a BP 
reading.

BP 
Reading

Immediately after a BP reading, the display shows the results of the 
reading if enabled. Blood pressure in mmHg followed by heart rate 
in beats per minute.

Battery

Indicates low battery voltage; BATTERIES NEED TO BE REPLACED.

Systolic/Diastolic

START/STOP

15

14

Содержание Opti 24-hour ABPM

Страница 1: ...Opti 24 hour ABPM User Guide...

Страница 2: ...14 Setting Up the System 15 Hardware Requirements 15 Software Requirements 15 Powering the Opti for use 15 Installing the Opti Insight Software 16 Conducting an Ambulatory Blood Pressure Study 17 Pro...

Страница 3: ...P Goodwin J 2004 Validation of the Opti oscillometric 24 hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices Blood Pr...

Страница 4: ...ndently programmable periods 5 10 15 20 30 45 60 90 and 120 minutes Size 120 x 70 x 32 mm Weight Approx 250 g including batteries Storage Conditions 20 c to 65 c 15 90 RH non condensing Data Connectio...

Страница 5: ...d potentially harmful injury to the patient WARNING Ensure batteries are inserted with the correct polarity Improper installation is a hazard WARNING Too frequent measurements can cause injury to the...

Страница 6: ...d manufacturer s declaration electromagnetic immunity The QRS Opti is intended for use in the electromagnetic environment specified below The customer or the user of the QRS Opti should assure that it...

Страница 7: ...xed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the QRS Opti device is used exceeds the applicable RF compliance level a...

Страница 8: ...age for two seconds 4 Three quick beeps 5 The number of BP readings in memory with flashing printer for three seconds 6 One long beep 7 Time flashing for twenty seconds The monitor is now ready to be...

Страница 9: ...onducting an Ambulatory Blood Pressure Study Communicating with the Opti To successfully complete an ABP study you need your computer to be able to communicate with your ABP monitor in order to progra...

Страница 10: ...allowed Time Intervals None 5 10 15 20 30 45 60 90 and 120 minute intervals between readings Fitting a patient with the Opti and OptiFit blood pressure cuff After you have successfully programmed the...

Страница 11: ...erify that the monitor operates correctly Finishing the study If you wish to finish the study before the patient returns instruct the patient to turn off the monitor by holding down the Start Stop but...

Страница 12: ...for servicing 99 Service Required EEPROM calibration data CRC failure Unit needs to be recalibrated Return for servicing In the event you are unable to rectify the error and need assistance please ca...

Страница 13: ...entative To return for service please send to QRS listed on Limited Warranty page 25 care of Customer Care Limited Warranty Opti Ambulatory BP Monitor QRS provides to the original purchaser the follow...

Страница 14: ......

Страница 15: ...Date 03 12 Advena Ltd 33 Bridge St Hereford HR4 9DQ UK 610001 00A 6901 E Fish Lake Road Suite 188 Maple Grove MN 55369 1 800 465 8408 1 763 559 8492 Fax 1 763 559 2961 info QRSdiagnostic com www QRSdi...

Отзывы: