background image

Programming the 

Opti

 for an ABP Study

To prepare the monitor for an ABP study, simply fill out an on-screen form to set the 
parameters for your patient to be programmed into the monitor.

1. Select the Program button on the 

toolbar or Program study under 
Monitor in the menu bar.

2. Enter the settings in the form (Fig. 

4). Fields are described below.

3. Programming begins when OK is 

clicked.

4. An indicator bar shows the 

progress as the data is 
transferred to the monitor and 
disappears when programming is 
successfully completed.

The test parameters can be adjusted as follows:

Patient name and ID:  For reporting and referencing data. 
Start study in 5 
minutes:

Check denotes that the study will start automatically after 
programming; unchecked denotes that the first push of the
Start/Stop button with unit powered on will start the study.

Time zone difference:  Adjust the monitor’s clock to the time zone that the patient 

is in relative to your time zone. 

Max Pressure: 

160 to 280 mmHg; suggested setting is 30 mmHg above 

the highest expected systolic BP. 

 

NOTE:

 The ABP monitor will not inflate to Max Pressure with each 

reading; it inflates to 30 mmHg above the previous systolic reading.

Keypad: 

Enabled will allow the patient to start readings. 

Display: 

Enabled will allow the patient to view the results 
immediately after a measurement. 

NOTE:

 Keypad and Display are always enabled for the first 30 

minutes of a study.

 

Intervals: 

Set intervals between programmed readings to Standard 

for +/- 5 minutes around selected times or Fixed for exact 
times. 5 and 10 minute intervals are always exact. 

Time Periods: 

Up to 3 allowed. 

Time Intervals: 

None, 5, 10, 15, 20, 30, 45, 60, 90, and 120-minute 
intervals between readings.

Fitting a patient with the 

Opti

 and 

OptiFit 

blood pressure cuff

After you have successfully programmed the 

Opti

 using 

Opti-Insight

, you may begin 

fitting the patient with the monitor and a blood pressure cuff.  Cuffs may be used on 
either arm.

1.  Choose the proper cuff size

  To determine the correct cuff size for your patient, wrap the cuff around the 

patient’s upper arm. Use the color-coded RANGE indicator on the inside of the 
cuff and the bold INDEX marker to check that the arm circumference falls within 
the cuff range. If the arm is within range, this cuff size is correct for your patient. 
If the measurement is outside the RANGE indicator, select a new cuff size as 
indicated by color.

 

IMPORTANT: 

Using an incorrect cuff size could 

result in erroneous and misleading blood pressure 
measurements.

2.  Apply the cuff 

  The cuff should be midway between the elbow and 

shoulder. Be sure the ARTERY indicator is over the 
patient’s brachial artery, between the bicep and tricep 
muscles. Wrap the cuff snugly around the patient’s 
upper arm.

3.  Connect the hoses 

  Connect the hoses from the cuff and monitor by twisting 

the fittings together until you hear a snap. Drape the hose over the patient’s 
shoulder, around the neck and across the opposite side of the body.

4.  Attach to patient 

  Insert the 

Opti

 into its pouch  

with the display visible. Attach 
the pouch to the patient using 
the belt.

5.  Begin BP reading 

  To verify proper monitor 

operation, ensure that the 
monitor is on and start a BP 
reading by pressing the Start/
Stop button. If problems occur, 
review the setup and fitting of  
the system or consult 
Troubleshooting for tips. 

Fig. 4: 

Program Monitor

Fig. 6: 

Application

Fig. 7: 

Proper Setup

Fig. 5: 

Application

19

18

Содержание Opti 24-hour ABPM

Страница 1: ...Opti 24 hour ABPM User Guide...

Страница 2: ...14 Setting Up the System 15 Hardware Requirements 15 Software Requirements 15 Powering the Opti for use 15 Installing the Opti Insight Software 16 Conducting an Ambulatory Blood Pressure Study 17 Pro...

Страница 3: ...P Goodwin J 2004 Validation of the Opti oscillometric 24 hour ambulatory blood pressure monitor according to the International Protocol for the validation of blood pressure measuring devices Blood Pr...

Страница 4: ...ndently programmable periods 5 10 15 20 30 45 60 90 and 120 minutes Size 120 x 70 x 32 mm Weight Approx 250 g including batteries Storage Conditions 20 c to 65 c 15 90 RH non condensing Data Connectio...

Страница 5: ...d potentially harmful injury to the patient WARNING Ensure batteries are inserted with the correct polarity Improper installation is a hazard WARNING Too frequent measurements can cause injury to the...

Страница 6: ...d manufacturer s declaration electromagnetic immunity The QRS Opti is intended for use in the electromagnetic environment specified below The customer or the user of the QRS Opti should assure that it...

Страница 7: ...xed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the QRS Opti device is used exceeds the applicable RF compliance level a...

Страница 8: ...age for two seconds 4 Three quick beeps 5 The number of BP readings in memory with flashing printer for three seconds 6 One long beep 7 Time flashing for twenty seconds The monitor is now ready to be...

Страница 9: ...onducting an Ambulatory Blood Pressure Study Communicating with the Opti To successfully complete an ABP study you need your computer to be able to communicate with your ABP monitor in order to progra...

Страница 10: ...allowed Time Intervals None 5 10 15 20 30 45 60 90 and 120 minute intervals between readings Fitting a patient with the Opti and OptiFit blood pressure cuff After you have successfully programmed the...

Страница 11: ...erify that the monitor operates correctly Finishing the study If you wish to finish the study before the patient returns instruct the patient to turn off the monitor by holding down the Start Stop but...

Страница 12: ...for servicing 99 Service Required EEPROM calibration data CRC failure Unit needs to be recalibrated Return for servicing In the event you are unable to rectify the error and need assistance please ca...

Страница 13: ...entative To return for service please send to QRS listed on Limited Warranty page 25 care of Customer Care Limited Warranty Opti Ambulatory BP Monitor QRS provides to the original purchaser the follow...

Страница 14: ......

Страница 15: ...Date 03 12 Advena Ltd 33 Bridge St Hereford HR4 9DQ UK 610001 00A 6901 E Fish Lake Road Suite 188 Maple Grove MN 55369 1 800 465 8408 1 763 559 8492 Fax 1 763 559 2961 info QRSdiagnostic com www QRSdi...

Отзывы: