Troubleshooting
Event
Code
Description in
Opti-Insight
Solution
1
Weak or no oscillometric signal
Check position of cuff, tighten cuff.
2
Artifact/Erratic Oscillometric Signal
Remain still during BP reading.
3
Exceeded retry count
(4 inflate attempts)
Remain still during BP reading.
4
Exceeded measurement time
Check air hose connections and
make certain cuff is tight.
85
Reading Aborted
(blocked valves or pneumatics)
Check air hose connections and
make certain air tubing is not
crimped.
86
Reading Aborted (user abort)
Push START/STOP button to restart
reading.
87
Reading Aborted
(inflate time-out or air leak)
Check air hose and cuff.
88
Reading Aborted (Safety time-out)
Retry reading, push START/STOP
button. If problems persist, return
unit for servicing.
89
Reading Aborted (cuff over-pressure)
Check for blocked or kinked air
hose.
90
Service Required
(power supply out-of-range or other
hardware problem)
Replace batteries. If problem
persists, return unit for servicing.
91
Service Required
(safety override fitted or autozero
out-of-range)
Retry by pushing START/STOP
button. If problems persist, return
unit for servicing.
97
Service Required
(transducer out-of-range)
Return for servicing.
98
Service Required
(A/D out-of-range)
Return for servicing.
99
Service Required
(EEPROM calibration data CRC
failure)
Unit needs to be recalibrated. Return
for servicing.
In the event you are unable to rectify the error and need assistance, please call our
service department: +1.800.465.8408 or +1.763.559.8492.
Maintaining and Cleaning the
Opti
After use, it is important to perform preventative maintenance to ensure the safe and
efficient operation of the monitor.
Cleaning after use
The
Opti
unit is not sterilizable. DO NOT immerse the monitor in any fluid, or
attempt to clean with any liquid detergents, cleaning agents, or solvents. You may
use a soft, damp cloth to remove dirt and dust from the monitor. If the unit does
become immersed in water, do not use; contact our service department.
You may use a mild disinfectant solution to clean the cuff, belt, and pouch.
Alternatively, you may also wash these items in a washing machine. Remove the
bladder from the cuff before machine washing. Wash these items using warm water
and a mild detergent; if needed, hang to dry.
Maintenance after use
Visually inspect cables, pneumatic hoses, and the monitor case for cracks, fraying, or
kinks. DO NOT use the monitor if there are any signs of damage. Please contact
our service department.
Maintenance
It is recommended that you check the accuracy of the
Opti
once every two years.
If needed, an authorized service center may need to recalibrate the pressure
transducers in the monitor.
23
22