QIAcuity Safety and Installation Guide 04/2021
81
4.
To clear an error, tap
Clear error
.
5.
If the error that occurred affects the tray(s), tap
More details
. To clear a tray-related error,
tap Clear error in the dialog. The dialog box contains five items that can be cleared for
each tray, such as motor and slot numbers (based on Insrument Version).
Note
: In QIAcuity Eight, the
More details
button is located in the Upper Tray and Lower Tray
panes. In the QIAcuity Four and QIAcuity One, the
More details
button is located in the Tray
pane.
6.
Restart the instrument. The instrument initializes and all modules are returned to their
home positions.
Note
: If the affected module is not working after you cleared the error and restarted the
instrument, contact QIAGEN Technical Services.
WARNING/
CAUTION
Risk of personal injury and material damage
[W1]
Only perform maintenance that is specifically described in this user
manual.