QIAcuity Safety and Installation Guide 04/2021
23
Grounding requirements
To protect operating personnel, the National Electrical Manufacturers’ Association (NEMA)
recommends that the QIAcuity be correctly grounded (earthed). The instrument is equipped
with a 3-conductor AC power cord that, when connected to an appropriate AC power outlet,
grounds (earths) the instrument. To preserve this protection feature, do not operate the
instrument from an AC power outlet that has no ground (earth) connection.
WARNING
Electrical hazard
Any interruption of the protective conductor (earth/ground lead)
inside or outside the instrument or disconnection of the protective
conductor terminal is likely to make the instrument dangerous.
Intentional interruption is prohibited.
Lethal voltages inside the instrument
When the instrument is connected to line power, terminals may
be live and opening covers or removing parts is likely to expose
live parts.
(W5)
Installation of AC power cord
The AC power cord connects to the socket located at the rear of the QIAcuity, and the other
end to the AC power outlet.