QIAcuity Safety and Installation Guide 04/2021
11
CAUTION
Damage to the instrument
[C2]
Only use QIAcuity-specific consumables with the QIAcuity. Do not use the
plates without applied top-seals. Damage caused by use of other
consumables will void your warranty.
CAUTION
Damage to the instrument
[C3]
Do not drop objects into the instrument when the plate tray is ejected.
WARNING
Risk of explosion
[W7]
The QIAcuity is intended for use with reagents and substances supplied with
QIAGEN kits or other than outlined in respective Information for Use. Use of
other reagents and substances may lead to fire or explosion.
CAUTION
Damage to the instrument
[C4]
Do not stack instruments and do not place items on top of the QIAcuity.
CAUTION
Damage to the instrument
[C5]
Do not lean against the touchscreen when it is pulled out.