MicroFlow-i HART Firmware Upgrade
1.
Please ensure that the sensor is connected to a PC via the HART modem.
Prior to reprogramming, you will need the HART firmware file, and the
COM port number of the modem.
2.
Establish a connection to the sensor.
3.
You will require Service access to Run HART Bootloader Control.
4.
Select Tools > Run HART Bootloader Control, which will then open the
Bootloader control interface window as seen in Fig.11.
5.
Once connected click File’, and choose the firmware file you are using to
upgrade the sensor.
6.
Click ‘Re-Program’, turn the sensor off and on again when prompted to do
so by the Bootloader control and then wait for the process to finish.
7.
When the ‘Program File OK now rebooting’ message is displayed, you can
close the HART Bootloader and reconnect to the PC software.
8.
Select the ‘Search’ button on the software, and the sensor will reconnect to
MicroFlow-i HART PC
with the new HART firmware installed.