Processing Parameters
These parameters relate to the detection and processing of the flow velocity signal.
Name
Parameter
No.
Options
Default
Notes
Gain
102
100 - 1600
120
Used to enter a fixed gain
(sensitivity) value, a higher
number means higher gain.
Quick response
Damping Factor
104
10 – 999
750
When P113=1, this damping
factor will be applied. When
P113=0, measurement will
only be damped by P109 the
main damping factor.
Cal Factor
108
1 – 500
100
In percentage terms, default
is 100, for 100%. 50%
would half and 200% would
double.
High/Main
Damping
109
0 - 28
0
This sets the damping factor
by adding to P104. A higher
number represents more
damping.
Step Response
Mode
113
0 = Off
1 = On
1
When turned off, no damping
by pass will be performed.
Low Damping
Persist
115
1 – 100
2
This is the number of
measurements that the sensor
acquires, before switching
into damping mode.
High Damping
Persist
116
1 - 100
1
This is the number of
measurements that the sensor
acquires, before switching
into Step Response Mode.
No Signal
Persist
117
1 – 100
2
This is the number of non-
measurement to persist
before the sensor sees a state
of no signal.
Signal Present
Persist
118
1 - 100
2
This is the number of
measurements that need to
persist for there to be a signal
present state.
Содержание MicroFlow-i
Страница 1: ...MICROFLOW INSTRUCTION MANUAL...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ...Page 6 EU Declaration of Conformity...
Страница 27: ...Page 21 Diagram 4 Diagram 3 Distance Distance Distance Distance Distance...
Страница 32: ...Page 26 This page is left blank intentionally...
Страница 34: ...Page 28 This page is left blank intentionally...