Using the Serial Interface
The RS232 serial interface is used to communicate between the FlowCERT
Lite and a PC using the optional Ultra PC and other associated Pulsar
software packages, to obtain information such as data logging and view
echo traces upload, download and save parameter files. In addition, it can
also be used to control or obtain information using a standard PC or other
computer base equipment. To do so, the settings for control are as follows:
baud rate
19,200
,
8 data bits
,
no parity
,
1 stop bit
.
The device should be connected as shown in
To use the device remotely, you need to
log on
to start, and
log off
when
finished. When
logged on
, FlowCERT Lite
will show ‘
Remote ON
’ on the
display, and “
Communicator OFF
” when
logged off
.
All commands should be followed by a carriage return
.
When logged on, the unit will respond either OK (or a value) if the command
is accepted, or NO if it is not.
To log on, send the command
/ACCESS:pppp where pppp is the passcode (P922).
To log off, send the command
/ACCESS:OFF
To read a parameter value, send the command
/Pxxx where xxx is the parameter you wish to read, and the FlowCERT will
respond with the parameter value.
To set a parameter, send the command
/Pxxx:yy where xxx is the parameter number, and yy is the value you wish to
set it to.