INCLINEO handbook
44
INCLINEO technical data
INCLINEO high precision inclinometer ALI 18.201
Measurement range
+10°
Resolution
0.0003° [1”]
Limits of error at
calibration [Ta =22°C]
0.005% full scale
0.03% read out
Limits of error at
measurement [Ta =22°C]
0.005% full scale
0.06% read out
8-hour zero-point drift
0.04% full scale
Digital Filter/Average
3rd order with 0.3 / 1 / 3 Hz options
Temperature range
Storage: -40°C to 85°C
Operation: -10°C to 60°C
Display
LCD display, 132 x 32 pixel with LED backlight
User interface
Three key operation
Wireless communication
Embedded RF module with LED indicator
External interface
RS-232 (serial) for computer and sensor
Connector for dial gauge
Power supply
2 AA batteries
Battery status indicator
3 LEDs
Data storage
Up to 999 measurements
Integrated RF module
Antenna – integrated
Transmission power – 20 dB max.
Complies with part 15 of the FCC rules. Operation is subject to following
conditions:
- Device must not cause harmful interference
- Device must accept any interference received, including interference
that may cause undesired operation
This transmitter is considered as mobile device and should not be used
closer than 20 cm from a human body.
Contained transmitter module — FCC ID: POOWML-C40 / CE: CE0678 !
USB RF dongle ALI 13.620
PC interface
USB 2.0 compatible
Temperature range
-40° C to 85° C
Dimensions
68 x 25 x 13 mm
Antenna
Integrated
Transmission power
19 dBm
Details
Complies with part 15 of the FCC rules. Operation is subject to following
conditions:
- Device must not cause harmful interference
- Device must accept any interference received, including interference
that may cause undesired operation
This transmitter is considered as mobile device and should not be used
closer than 20 cm from a human body.
Contained transmitter module — FCC ID: QOQWT41 /
IC: 5123A-BGTWT41