8. With INCLINEO still in its original position, mark the current location to enable
INCLINEO be placed at exactly the same position for the reversal measurement.
9. Now rotate INCLINEO through 180°.
10. Align INCLINEO into place then press
ENTER
to start second measurement.
11. After measurement has stabilized “ok” appears on the display. The system has
now been set for absolute measurements.
12. Press
ENTER
to confirm setting the level condition.
13. With the submenu item “save” displayed, press
ENTER
to save the set-up. Readings
taken henceforth are absolute values.
For the second reading,
INCLINEO is rotated
through 180° and placed
at exactly the same
position as in the first
reading
BAT T
INCLINEO
®
ENTER
MENU
HOLD
COMPUTER
SENSOR
GAUGE
Type
www.pruftechnik.co
m
Made in Ger
m
any
s.No.
HW
BATTERY
ON
OFF
1
2
Turn INCLINEO and press
g
g
ENTER to
m
easure
8
9
13