44
FUNCTIONAL TESTS
Pass
Fail
Front Panel Test
Self Test
Drive Voltage
Resistance Test
24.9
400
6.4 k
100 k
Capacitance Test
FREQUENCY ACCURACY
Test Frequency
Minimum
Actual
Maximum
100 Hz
99.99 Hz
100.01 Hz
120 Hz
119.99 Hz
120.01 Hz
1 kHz
999.90 Hz
1000.1 Hz
10 kHz
9999.0 Hz
10001.0 Hz
100 kHz
99990.0 Hz
100010.0 Hz
AMPLITUDE ACCURACY
Amplitude
Frequency
Minimum
Actual
Maximum
1.0 Vrms
1 kHz
0.98 Vrms
1.02 Vrms
1.0 Vrms
100 Hz
0.98 Vrms
1.02 Vrms
1.0 Vrms
120 Hz
0.98 Vrms
1.02 Vrms
1.0 Vrms
10 kHz
0.98 Vrms
1.02 Vrms
1.0 Vrms
100 kHz
0.98 Vrms
1.02 Vrms
0.25 Vrms
1 kHz
0.245 Vrms
0.255 Vrms
0.1 Vrms
1 kHz
0.098 Vrms
0.102 Vrms
Internal Bias
1.96 VDC
2.04 VDC
Содержание 9216A
Страница 1: ......
Страница 2: ...PROTEK 9216A Digital LCR Meter User s Manual Volume 1 Basic Operation Rev 3 0 ...
Страница 74: ...43 9216A TEST PERFORMANCE RECORD Serial Date Tested by Equipment used ...
Страница 77: ...46 ...