14
Table 2-7 Parameters for Calculating K
l
and K
h
for Resistive Measurements
R
l
K
l
Z
m
R
h
K
h
Z
m
Frequency
R3
R2
R1
R0
R3
R2
R1
R0
100, 120, 1 kHz
1 m
0.02
0.2
4
400 k
6.5 M
100 M
2 G
10 kHz
1 m
0.02
0.2
4
400 k
6.5 M
100 M
1.5 G
100 kHz
4 m
0.03
0.4
---
200 k
3 M
50 M
---
Table 2-8 Parameters for Calculating K
l
and K
h
for Inductive Measurements
L
l
K
l
L
m
L
h
K
h
L
m
Frequency
R3
R2
R1
R0
R3
R2
R1
R0
100, 120 Hz
1
H
20
H
300
H
5 mH
630 H
10 kH
160 kH
2.6 MH
1 kHz
1
H
20
H
300
H
5 mH
630 H
10 kH
160 kH
2.6 MH
10 kHz
1
H
20
H
300
H
5 mH
630 H
10 kH
160 kH
2.6 MH
100 kHz
1
H
20
H
300
H
5 mH
630 H
10 kH
160 kH
2.6 MH
Table 2-9 Parameters for Calculating K
l
and K
h
for Capacitive Measurements
C
l
K
l
C
m
C
h
K
h
C
m
Frequency
R3
R2
R1
R0
R3
R2
R1
R0
100, 120 Hz
4 nF
240 pF
16 pF
2 pF
1600 mF
80 mF
6.7 mF
400
F
1 kHz
400 pF
24 pF
1.6 pF
0.1 pF
160 mF
8 mF
670
F
40
F
10 kHz
40 pF
2.4 pF
0.16 pF
0.01 pF
16 mF
800
F
40
F
2.6
F
100 kHz
8 pF
2.4 pF
0.02 pF
---
200
F
80
F
4
F
---
Verification of Meter Performance
The performance verification procedures in this section test and verify the performance of
9216A and compare it to the specifications listed in Volume 1 of this manual. The first set
of tests verifies the basic functionality of the unit. The second set of tests verifies the criti-
cal specifications of the 9216A. The results of each section can be recorded on the test
sheet located at the end of this manual.
Functional Tests
These simple tests verify the basic functionality of the 9216A. They are not intended to
verify the accuracy of the unit.