30
BINNING
BCL
BING(?) {
i
}
BLIM(?)
i
j
{
x
}
BNOM(?)
i
{
x
}
Clears nominal values and limits for all bins. All bins are closed.
Set (query) binning to enabled (
i
1) or disabled (
i
0).
Set (query) upper (
i
0) or lower (
i
1) limit of bin
j
(0-7) to
x
%.
Set nominal value of bin
i
to
x
SETUP CONTROL
*IDN?
*OPC(?)
*RCL {
i
}
*RST
*SAV{
i
}
*WAI
Returns the 9216A identification string.
Set bit in Standard Event Status byte when measurement is complete.
Recall setting
i
.
Reset unit to default configuration.
Save current setup as setting
i
.
Wait until all measurements are completed before proceeding.
STATUS
*CLS
*ESE(?) {
i
}
*ESR? {
i
}
*PSC(?) {
i
}
*SRE(?) {
i
}
*STB? {
i
}
SENA(?) {
i
}
STAT? {
i
}
Clear all status registers.
Set (query) the Standard Event Status Byte Enable register to value
i
(0-255).
Query Standard Status byte. If
i
is included, only bit
i
is queried.
Set (query) power-on status clear bit to clear (
i
1) or maintain (
i
0) status val-
ues.
Set (query) the Serial Poll Enable register to value
i
(0-255).
Query Serial Poll status byte. If
i
is included, only bit
i
is queried.
Set (query) LCR Status Enable register to value
i
(0-255).
Query LCR Status byte. If
i
is included, only bit
i
is queried.
Status Displays and Error Messages
Adjacent to the right alphanumeric displays, the 9216A has three front panel status LEDs
that provide positive indication of commands and communications to help verify opera-
tions during the development of control programs. The ACT LED flashes whenever a
character is received or sent over either interface. The ERR LED flashes when an error
has been detected, such as an illegal command, or parameter out of range. The REM
LED is lit whenever the 9216A is in a remote state (front panel locked out).
The 9216A reports two types of errors that may occur during command execution:
command errors and execution errors. Command errors are errors in the command syn-
tax. For example, unrecognized commands, illegal queries, lack of terminators, and non-
numeric arguments are examples of command errors. Execution errors are errors that
occur during the execution of syntactically correct commands. For example, out of range
parameters and commands that are illegal for a particular mode of operation are classi-
fied as execution errors.
The NO COMMAND bit is a bit in the serial poll register that indicates that there no
commands waiting to be executed in the input queue. This bit is reset when a complete
command is received in the input queue and is set when all of the commands in the
queue have been executed. This bit is useful in determining when all of the commands
sent to the 9216A have been executed. This is convenient because some commands,