41
Bias Error
7
The 9216A internal DC bias source failed its test.
Cal Error
4
The RAM calibration data has become corrupt. The factory
values will be reloaded from ROM. This message is not a
problem unless it occurs frequently, which could indicate a
problem with the battery backup circuits.
Code Err XX
2
The 9216A ROM has a checksum error. XX is the checksum
value.
CPU Error
1
The 9216A has detected a problem in its CPU.
Data Error
3
CPU RAM failed a read/write test.
Det Error
6
The square wave multiplier failed its DC rejection test.
Drv Error
i
7
The output drive circuitry failed its test.
i
is an error code indi-
cating the failure point.
i
error
0
100 Hz amplitude failure
1
120 Hz amplitude failure
2
1 kHz amplitude failure
3
10 kHz amplitude failure
4
100 kHz amplitude failure
5
0.25 V amplitude attenuator failure
6
0.1 V attenuator failure
Freq Error
i
5
The frequency clock generator failed its test.
i
is an error code
indicating the failure point.
i
error
0
100 Hz failure
1
120 Hz failure
2
1 kHz failure
3
10 kHz failure
4
100 kHz failure
Gain Error
8
The instrumentation amplifier failed its gain test.
i
is an error
code indicating the failure point.
i
error
0
the
2 gain failed
1
the
4 gain failed
2
the
8 gain failed
3
the
20 gain failed
Out Error
9
The source impedance selection circuitry failed its test. This
error can occur if a part is in the fixture during the test. i is an
error code indicating the failure point.
i
error