19
Resistance Accuracy Table
resistance
range
conditions
Tol R tol.
10.0
25.0
25.0
25.0
100.0
3
3
3
3
3
1 kHz series
1 kHz series
10 kHz series
100 kHz series
1 kHz series
9.9930 - 10.007
24.983 - 25.018
24.970 - 25.030
24.945 - 25.055
99.930 - 100.07
100.0
400.0
400.0
400.0
1.600 k
2
2
2
2
2
1 kHz series
1 kHz series
10 kHz series
100 kHz series
1 kHz parallel
99.930 - 100.07
399.72 - 400.28
399.52 - 400.48
399.12 - 400.88
1.5989 k - 1.6011 k
1.6000 k
6.4000 k
6.4000 k
6.4000 k
25.0000 k
1
1
1
1
1
1 kHz parallel
1 kHz parallel
10 kHz parallel
100 kHz parallel
1 kHz parallel
1.5989 k - 1.601 k
6.3955 k - 6.4045 k
6.3923 k - 6.4077 k
6.3859 k - 6.4141 k
24.983 k - 25.018 k
25.0000 k
100.0000 k
100.0000 k
400.0000 k
0
0
0
0
1 kHz parallel
1 kHz parallel
10 kHz parallel
1 kHz parallel
24.983 k - 25.018 k
99.930 k -100.07 k
99.880 k -100.12 k
399.72 k - 400.28 k
* if the Q value is greater than 0.1, the allowed tolerance must be multiplied by (1+Q)
Capacitance Accuracy
1. Connect the adapter ground leads of IH and IL to ground terminal of the decade
capacitor. Connect the IH and PH leads to the ―
+
‖
terminal of the decade capaci-
tor. Connect the IL and PL leads to the "
"
terminal of the decade capacitor box.
Set the 9216A to its default conditions by pressing the key sequence
RCL 0 EN-
TER
, then select the R+Q measurement mode and the 1 kHz test frequency.
2. Unplug the IH, PH leads from "+" terminal of the decade capacitor box and plug
them into the "
" terminal of the decade capacitor box. Perform short circuit cali-
bration. Unplug the IH, PH leads from "
" terminal of the decade capacitor box,
then perform open circuit calibration.
3. Connect the IH, PH leads to the "+" terminal of the decade capacitor box. Set the
capacitance to zero.
4. Note the capacitance value the 9216A measures. This is the capacitance of the
switch contacts and connectors. Enter this as a relative value in the entry display.
Set the 9216A to the DEV display mode.
5. Set the decade capacitor and the 9216A according to the values in the Capaci-
tance Accuracy Table below. Verify that the readings fall within the acceptable
values. Record the results.