22
Table 2-11 Organization of Floating Point Calbytes
Calbyte Number
Name
Meaning
0
Frequency Correction
Deviation from nominal, in ppm
1
–40
Reference Resistor Values
Impedance values organized in real and imaginary
pairs
41
–80
Open Circuit Values
Admittance values organized as real and imagi-
nary pairs
81
–120
Short Circuit Values
Impedance values organized in real and imaginary
pairs
Necessary Equipment and Conditions
To perform the amplitude, frequency, or standard resistor calibration, the following
equipment is needed:
A time interval counter, with an accuracy of 1 ns or better.
An AC/DC voltmeter, with 5 ½ -digit DC accuracy and true RMS AC measure-
ments to 100 kHz.
Standard resistors, with resistance known to 0.01% and Q accuracy known to 25
ppm. The resistor values needed are 95.300 k
(R0), 5.970 k
(R1), 374.0
(R2), and 25.10
(R4).
A fixture-to-BNC adapter.
The conditions for calibration include a warm-up time of at least 30 minutes and an ambi-
ent temperature of 23
C
2
C.
Amplitude Calibration
This procedure calibrates the output amplitude at the different frequencies and output
voltages. To adjust the calbytes, press the
CAL
key until the message "cl
I
" appears in
the display, where
I
is an integer. The present value of that calbyte will appear in the right
display. The
and
frequency keys step through the different amplitude calbytes. New
values are entered using the numeric entry keys.
1. Connect the AC/DC voltmeter across the two sides of the fixture. A small piece of
wire inserted in each side of the fixture is a convenient way to connect the DVM.
Do not connect either end to ground. Set the meter to AC volts, auto ranging.
Set the 9216A to its default conditions by pressing the keys,
RCL
0
ENTER
. Set
the unit to constant voltage mode.
2. Measure the amplitude and frequency for amplitude calbyte 0 (0.10 V and 100
Hz). If the value is not within 2% of the nominal value, enter the new calbyte us-
ing the formula:
byte
CurrentCal
Vmeas
om
Vn
NewCalbyte
, rounded to the nearest integer