Page 23
6
Measurement File
All measurement files are stored in the internal memory and can be
downloaded via USB. When the ZGM 1130 is connected to USB
choose Download and the device will appear as a flash drive on the
Computer. Measurement files are comma separated (CSV) and can be
viewed with any notepad or spreadsheet program.
Each measurement file can contain up to 1000 single measurements. These are
structured as shown below. The measurement file starts with header information
about the device and the supported angles.
Measurement files can be exchanged between ZGM 1130´s. However the
header information will not change, new measurements will just be added at
the bottom of the file
Each measurement is structured in the following 12 rows.
No.
Ascending number of a single measurement
20°
Measurement value of the 20° angle in GU or %
60°
Measurement value of the 60° angle in GU or %
85°
Measurement value of the 85° angle in GU or %
H
Measurement value of reflection haze in HU
Date
Date the measurement has been taken [DD.MM.YYYY]
Time
Time the measurement has been taken [hh:mm:ss]
Sequence
Ascending sequence number. This number is incremented
whenever a predefined number of sequence measurements have
been done or when the measure mode has been changed.
IsOnStandard
A flag which indicates if the measurement has been taken on the
calibration standard or on the sample.
‘1’ = on the calibration standard
‘0’ = not on the calibration standard
Mode
Indicates the measure mode. Can be “Standard”, “Sequence” or
“Scan”
Comment
not used in the ZGM 1130
SampleName
not used in the ZGM 1130
Note: Please be aware that the statistic values are not stored in the
measurement file. However, statistic values can easily be calculated with
any spreadsheet program.