Page 19
Info
- General information about the device
will be shown here e.g. when the factory
calibration expires and a recalibration from
the manufacturer is recommended.
4.8.4 Modes
The ZGM 1130 supports three different measure modes. Standard and Sequence
mode are used to trigger one measurement
at a time. Whereas Scan mode measures
continuously with high speed as long as the
measure button is pressed. This enables to
measure an average value over a much
larger area than just the measuring area
itself.
In
Standard
mode,
only single
measurements are taken and the statistic values are reset when the mode has
been changed or the same mode has been chosen again.
In
Sequence
mode,
the number of single measurements which should be made
and lead to an average value can be predetermined. The statistic values are reset
after this predefined number of single measurements, when the mode has been
changed or the same mode has been chosen again.
In
Scan
mode,
the ZGM 1130 measures with high speed as long as the measuring
button is pressed. All single measurements completed during that time lead to one
single average value which will be added to the statistic display. Treated as one
measurement. The statistic values are reset when the mode has been changed or
the same mode has been chosen again.