This function allows to delete whole measurement files from the ZGM 1130. Choose
the measurement file to delete from the list and confirm or decline. If the current
measurement file will be deleted, a new one named DEFAULT will be created
automatically.
4.8.2 Calibrate
It is necessary to calibrate the device from time to time. This is may be required
particularly if ambient conditions have been changed since the last calibration. It is
recommended to calibrate the device at least every 2 days.
This function will automatically calibrate all
available geometries against a calibration
standard. The current calibration values are
shown on the entry window of a calibration.
These can be changed within the setup
menu (e.g. for calibration on a second
standard). When pressing the Calibration
button the ZGM 1130 moves on with the
calibration. If the device is not placed on the
calibration standard a message is shown on
the display.
After the device has performed a calibration, a summary will be shown indicating
which angles exceeding a certain tolerance.
Confirm the calibration with
or discard it with
.
This angle has been calibrated successfully without any error.
This angle has been calibrated successfully without any error but the device
has not been placed on its calibration standard.
An error occurred during calibration. Please consider the message shown.