Lumen 200Pro Standard Shutter Commands.
The Lumen 200Pro Shutter accepts the standard commands for a Filter Wheel, Position 1-10
provide a standard range of light output from 0-100%.
Command Arguments Response
(All end
with <cr>)
Description
7
w, f
if f = F the
current
filter else R
If no wheel
is fitted
E,17 will be
returned.
If f is a number move filter wheel w to filter position f.
if f is a ‘N’ move filter wheel w to next filter.
If f is a ‘P’ move filter wheel w to previous filter.
If f is a ‘F’ report current filter on filter wheel w.
if f is a ‘H’ performs a home routine.
If f is ‘A’ wheel will auto home on controller startup
If f is ‘D’ wheel will NOT auto home on startup (default)
FILTER
w
Text string Prints information about filter wheel w. The information end is
always a line saying END This allows for the addition of extra
fields of information without effecting application software.
Users should always read lines in until the END is seen to keep
in order to maintain compatibility.
Example
FILTER_1 = LLG_SHUTTER
TYPE = 3
PULSES PER REV = 262500
FILTERS PER WHEEL = 10
OFFSET = 223500
HOME AT STARTUP = FALSE
END
FPW
w
n
Reports the number n of filters on wheel ‘w’.
SAF
w
a
Report the current filter wheel w acceleration setting.
SAF
w, a
0
Sets the current filter wheel w, acceleration to a. Range is
4 to 100