40
P rior Sc ie ntific
5.2.4
Connecting your Lumen 200Pro to Prior PCI controller
The Lumen200Pro requires two filter wheel connections to a Prior PCI system.
Connecting your system:
1.
Switch off the PCI controller/computer.
2.
Connect the Attenuator/shutter to “FILTER 1” using the cables provided.
a.
Tighten the screws to fix the cables in position.
3.
Connect the Filter Wheel to “FILTER 2” using the cables provided.
a.
Tighten the screws to fix the cables in position.
4.
Connect lamp ON/OFF to “Shutter 1” connection using cable provided, (Optional).
a.
If connected lamp default to OFF.
5.
Switch on the controller and computer, your Lumen200Pro is now ready to accept
commands.
a.
It is not necessary to switch the Lumen200Pro power on to control the Shutter and
Filter Wheel, these are powered by the attached controller.