32
Save Button
Once the calibrations are complete, it is important to press the save button to set the center position to the correct
value. When performing label analysis and if the boxes are not aligned around the barcodes, this will be because of
not pressing the save button.
Tap Test
When unit is tight, and unit is still ON, tap on ODV unit and verify the dial gages all return into the green area.