Page 7
2.2. Specifications
- volt and mA Inputs
Input Ranges
Resolution
Accuracy
Remarks
volt
0 to 11 V
11 V to 30 V
0.0001 V
0.0001 V
0.02 % FS
0.02 % FS
R
input
> 1 M
mA
0 to 24.5 mA
0.0001 mA
0.02 % FS
R
input
< 65
- volt and mA Outputs
Input Ranges
Resolution
Accuracy
Remarks
volt
0 to 11 V
0.0001 V
0.02 % FS R
output
< 0.3
mA
0 to 22 mA
0.0001 mA
0.02 % FS
R
max
= 450
2-Wire Transmitter
(XTR)
4 mA to 22 mA
0.0001 mA
0.02 % FS
V
max
= 30 V
- Probe
Probe Range
Resolution
Accuracy*
Remarks
Pt-100 -200 °C to 850 °C /
-328 °F to 562 °F
0.01 °C /
0.01 °F
0.1 °C
0.2 °F
IEC-60751
(*)The accuracy is related only to PC-507-IS input.
presys
Содержание PC-507-IS
Страница 1: ...PC 507 IS Pressure Calibrator TECHNICAL MANUAL presys Instruments p r e s y s ...
Страница 10: ...Page 9 2 3 Order Code p r e s y s ...
Страница 13: ...Page 12 3 Operation 3 1 Parts Identification Front Panel Fig 01 Front Panel p r e s y s ...
Страница 14: ...Page 13 Left side view Panel Right side view Panel Fig 02 Side Panels p r e s y s ...