Page 51
Probe Adjustment.
First identify the connector pins of
Probe
input according to the figure below:
Fig. 27
– Probe Adjustment
Probe
adjustment is performed in two steps:
a) mV Signal Application:
mV signal
Terminals
1
st
point
2
nd
point
mV signal
V_2
RTD2(+)
V GND
100.000 mV
120.000 mV
V_1
RTD2(+)
V GND
120.000 mV
600.000 mV