Page 54
4.2. Output Adjustment
Select the corresponding mnemonic, choose the setpoint explained as
follows, measure the signal generated by the calibrator and store this value as
detailed in the following tables.
For adjustment of outputs, the display shows three values:
SP = 50,000%
(1)
49,999
(2)
5,00000
(3)
First Field (1) is the value of the setpoint as a percentage (%) of the
output range required by the user and it is selected by pressing key "0" (zero).
Second Field (2) is the value measured by the calibrator expressed in
percentage (%) of the output range. Before providing the 1
st
, 2
nd
and 3
rd
adjustment
points, one must wait until this value stabilizes.
Third Field (3) is the value entered by the user after the output has been
measured and the values corresponding to the two setpoints have been stored: 1
st
point (key 1), 2
nd
point (key 2) and 3
rd
point (key 3).
V output
1
st
point
2
nd
point
3
rd
point
Single range
SP = 10.000%
SP = 60.000%
SP = 80.000%
mA output
1
st
point
2
nd
point
3
rd
point
Single range
SP = 10.000%
SP = 60.000%
SP = 80.000%
presys
Содержание PC-507-IS
Страница 1: ...PC 507 IS Pressure Calibrator TECHNICAL MANUAL presys Instruments p r e s y s ...
Страница 10: ...Page 9 2 3 Order Code p r e s y s ...
Страница 13: ...Page 12 3 Operation 3 1 Parts Identification Front Panel Fig 01 Front Panel p r e s y s ...
Страница 14: ...Page 13 Left side view Panel Right side view Panel Fig 02 Side Panels p r e s y s ...
Страница 31: ...Page 30 a 4 wire Transmitter Calibration Fig 13 4 WIRE Transmitter Calibration p r e s y s ...