Page 35
Connect the pressure switch relay output to the contact input of the Calibrator,
V IN and mA OUT (+) terminals, and make the pneumatic connections illustrated in
the figure below.
Select
SW
(switch) from the input menu of the calibrator and
Pressure
from
the output menu.
Vary the pressure manually. The contact input will track the output pressure
reading until the relay position changes. At this time, the relay input reading will
'freeze' with the
LOCK
message in the upper line of the display. The value next to the
LOCK
message is the pressure switch alarm setpoint.
To unlock the input, press the arrow key .
Fig. 20
– Pressure Switch
presys