Page 52
b) Standard Resistors Application:
Connect a decade box or standard resistors to the
Probe
connector, at
positions RTD1, RTD2, RTD3 and RTD4, as shown above.
Resistors
1
st
point
2
nd
point
R_2
20.000
50.000
R_1
100.000
500.000
Pressure Adjustment
By selecting
Press
(ENTER), it shows the sub-menu:
V_1
V_2
P1
P2
Pressure
input adjustment is only completed after calibrating the options:
V_1
,
V_2
,
P1
and
P2
. Note that
P1
and
P2
refer to two pressure sensors chosen
from
P
option in
CONF
menu
The
V_1
and
V_2
adjustment is performed only at the factory.
The calibrations of the sensors associated to mnemonics
P1
and
P2
consist of
the application of two pressure values close to zero and full scale (FS) of the
pressure sensor. Store the values of the 1
st
and 2
nd
calibration points by means of
keys 1 (1
st
point) and 2 (2
nd
point).
presys