Page 19
lower line of the display (chosen in
OUT
). Note that the units are distributed in
several lines. In order to move the cursor through the units use the keys:
,
,
and . After selecting the unit, press ENTER.
The available units are written below between parentheses beside the
corresponding mnemonics presented in the calibrator display:
PSI
(psi)
ATM
(atm)
KG
(kgf/cm
2
)
INW
(inH
2
O)
MW
(mH
2
O)
CMW
(cmH
2
O)
MMW
(mmH
2
O)
INH
(inHg)
CMH
(cmHg)
MMH
(mmHg)
BAR
(bar)
MBA
(mbar)
KPA
(kPa)
TOR
(torr)
PA
(Pa)
PARAM
- configures the calibration parameters of the pressure sensors
provided by the manufacturer.
DT -
updates the date and time for the calibrator. Thus, when it performs a
calibration within the communication option, calibration data is recorded together with
their date and time of occurrence.
Whenever the calibrator is turned off, the date and time are not updated any
longer. Thus, if you want them to be recorded with the calibration, you should update
them. Use the vertical arrow keys
and
to change the value which is blinking and
the horizontal arrow keys and to go to another value. The ENTER key confirms
the last selection
presys
Содержание PC-507-IS
Страница 1: ...PC 507 IS Pressure Calibrator TECHNICAL MANUAL presys Instruments p r e s y s ...
Страница 10: ...Page 9 2 3 Order Code p r e s y s ...
Страница 13: ...Page 12 3 Operation 3 1 Parts Identification Front Panel Fig 01 Front Panel p r e s y s ...
Страница 14: ...Page 13 Left side view Panel Right side view Panel Fig 02 Side Panels p r e s y s ...
Страница 31: ...Page 30 a 4 wire Transmitter Calibration Fig 13 4 WIRE Transmitter Calibration p r e s y s ...