39
MMD
<0,?>
<0,?>
Measuring MoDe
Only argument <0>, chromatic mode supported
NOP
<1..4>
<1..4>
Number Of Peaks
1(default)-MaxNOP.
The MaxNOP value is a characteristic of the device and
can be read back by the IDE command.
In confocal mode, if less than NOP peaks are detected, all
thickness signals will be invalidated because peak
identification is not possible.
OPD
<?>
<Operation time in seconds, number
of powerups>
OPeration Data
Query command, must contain the query arg.
First parameter 2: Operation time in seconds
First parameter 3: number of powerups
PSM
<10 .. 90>
<10 .. 90>
Peak Separation Minimum
The fraction in % of the height of the smaller peak that
the signal must go below in order to separate 2 peaks.
RST
ReSeT device
SCA
<?>
<Scale value>
SCAle
Query full scale value in micrometers of the currently
selected probe.
SEN
<0 .. 7, ?>
<optical probe Index>
Select SENsor (probe)
0(default) - 7 (depending on the existence of the calibration
tables)
See also: SENX
SENX
<?>
<(probe index), SNr: (serial number
of probe), Range: (measuring
range)um>
Query details of SENsor (probe)
Without parameter: returns a string with serial number
and measuring range of the currently selected probe in
the following format:
(probe index), SNr: (serial number of probe), Range:
(measuring range)um
When used with enum as string parameter, the device
enumerates all available calibrated probes in one string,
separated by ;
See also: SEN
SFD
-
-
Set Factory Defaults
Without parameter: reset the normal operation
parameters.
With parameter: reset also the communication
parameters (IP).
The parameter set is NOT automatically stored in
nonvolatile memory. To do so, please issue the SSU
command!
SHZ
<30 .. 4000, ?>
<x>Hz
x meaning the exact sample rate in
Hz in floating point format
Set measurement cycle frequency in HZ
30 - 4000(default)
See also: SRA
SODX
<0..17> <0..17>
…<0..17>
(max 16 times)
or <?>
<0..17> <0..17> … <0..17> (max 16
times)
Set output data extended
Set the content of the output telegram
Содержание CHRocodile C
Страница 1: ...1 CHRocodile C Compact sensor for non contact distance and thickness measurement Operation Manual...
Страница 3: ...3 Version Control Version Manual Date Type of Change 1 0 0 0 2016 09 20 Original edition...
Страница 13: ...13 Fig 2 2 Chromatic Confocal Imaging principle point sensor...
Страница 57: ...64 Appendix 2 Mechanical Plans 1 Optical probe mechanical plans b Probe 1mm a Probe 0 2mm...
Страница 58: ...65 Fig A2 1 Optical Head Mechanical plans a 0 2mm b 1mm c 4mm d 10mm c Probe 4mm d Probe 10mm...
Страница 59: ...66 2 CHRocodile C unit mechanical plans Fig A2 2 CHRocodile C sensor unit mechanical plans...