53
2.7
NOP Command
Short description:
This command allows to set the number of peak to evaluate.
WARNING
:
In confocal mode, if less than NOP peaks are detected, all thicknesses signals will
be invalidated because peak identification is not possible.
Command syntax:
$NOP <I>
Param: <I> is Number of Peak from 1 to 4
Command
Description
$NOP 3
Response: $NOP 3[CR]ready[CR/LF].
Write Number of Peak (3)
$NOP ?
Response: $NOP ? 3ready[CR/LF].
Read Number of Peak
2.8
SCA Command
Short description:
The command Scale allows to query of Full Scale in micrometers.
A distance value of 32768 on the serial interface would mean a distance in (Full Scale)
micrometers. To convert the integer distance value (d) received from the serial interface to a
value in micrometers (D), use the formula:
D[µm] = d[integer] / 32768 * Full Scale.
Command syntax:
$SCA
Command
Description
$SCA ?
Response: $SCA ? 3320ready[CR/LF].
Read Full Scale
2.9
SEN / SENX Command
Short description:
The $SEN command is used to activate the chromatic—confocal calibration related to a specific
sensor head. With the single channel device, up to sixteen calibrations can be uploaded and
stored in the device. With the Chrocodile C device, up to eight calibrations are supported.
Содержание CHRocodile C
Страница 1: ...1 CHRocodile C Compact sensor for non contact distance and thickness measurement Operation Manual...
Страница 3: ...3 Version Control Version Manual Date Type of Change 1 0 0 0 2016 09 20 Original edition...
Страница 13: ...13 Fig 2 2 Chromatic Confocal Imaging principle point sensor...
Страница 57: ...64 Appendix 2 Mechanical Plans 1 Optical probe mechanical plans b Probe 1mm a Probe 0 2mm...
Страница 58: ...65 Fig A2 1 Optical Head Mechanical plans a 0 2mm b 1mm c 4mm d 10mm c Probe 4mm d Probe 10mm...
Страница 59: ...66 2 CHRocodile C unit mechanical plans Fig A2 2 CHRocodile C sensor unit mechanical plans...