69
-
Check if the target surface is normal to the optical probe axis. The maximum angle between
target surface and optical axis is given by the maximum measurable slope (Cf. section 2.8).
-
Check if the distance is transmitted ($SODX 16640 or SODX 256),
-
Check that a calibration table is uploaded ($SCA?)
-
Then, check and store the raw signal (this file could be demanded by PRECITEC technical
support team). The raw signal should show one peak.
1.4 Thickness measurement:
The target is positioned in front of the sensor, and no thickness measurement is collected for all
channels.
-
Check that the target thickness is inside minimal and maximal measuring thickness range,
-
Check the target transparency,
-
Set the measuring frequency to minimum ($SHZ100).
-
Set the LED intensity to maximum ($LAI100).
-
Check if a white light spot is emitted from the optical probe and if it is focused on the target.
-
Check if the two surfaces of the target are inside the measuring range of the optical probe (Cf.
section 2.8).
-
Check if the target surface is normal to the optical probe axis. The maximum angle between
target surface and optical axis is given by the maximum measurable slope (Cf. section 2.8).
-
Check if the two distance peaks are transmitted ($SODX 16640 16648),
-
Then, check and store the raw signal (this file could be demanded by PRECITEC technical
support team). The raw signal should show two peaks.
If after reading the previous section you didn’t succeed in resolving your problem, please contact your
vendor for technical support. In order to be more efficient, we recommend to fill the following technical
support datasheet and if necessary prepare the files which are demanded in previous section.
Содержание CHRocodile C
Страница 1: ...1 CHRocodile C Compact sensor for non contact distance and thickness measurement Operation Manual...
Страница 3: ...3 Version Control Version Manual Date Type of Change 1 0 0 0 2016 09 20 Original edition...
Страница 13: ...13 Fig 2 2 Chromatic Confocal Imaging principle point sensor...
Страница 57: ...64 Appendix 2 Mechanical Plans 1 Optical probe mechanical plans b Probe 1mm a Probe 0 2mm...
Страница 58: ...65 Fig A2 1 Optical Head Mechanical plans a 0 2mm b 1mm c 4mm d 10mm c Probe 4mm d Probe 10mm...
Страница 59: ...66 2 CHRocodile C unit mechanical plans Fig A2 2 CHRocodile C sensor unit mechanical plans...