Connections and Interfaces
All of the connection ports are located at the rear of the CHRocodile C (Cf. section 2.6 Connections
and Interfaces):
The power supply jack,
The RJ45 standard connector for Ethernet communication,
The multipoint connector interface (9 pins) for trigger input / output and RS422/RS232 serial
communication,
3.1.1
CHRocodile C Stand Alone device:
The device can be used as a stand-alone device in order to perform selective distance or thickness
measurements. In this condition, only the power supply and Ethernet connector or multipoint
connector are useful.
Power supply
The CHRocodile C has two pluggable screw terminal for power supply with 24V
DC
+/-10%.
Connect the set of power cable supply associated to the Power Supply Adapter (100-240VAC to
24V
DC
+/-10%) delivered with the CHRocodile C unit.
Ethernet connector
Connect the isolated RJ45 standard connector from the CHRocodile C unit to an Ethernet network
(PC). Use shielded cable for the data port connection (minimum category 5 cable).
The default CHRocodile C IP address is:
192.168.170.4
Configure the PC Ethernet port to the following address:
192.168.170.X (X≠4)
The CHRocodile C can also be configured to use auto-configuration of the IP-Address through
a DHCP server.
To configure the Ethernet port of your PC,
you must open the ‘
Network connection properties’
menu.
After selecting the right Ethernet card (connected to the sensor), click on ‘
network protocol (TCP/IPv4)’
and click ‘
Properties’.
Set the IP address of the PC and the mask. For a standard use the mask should
be set to 255.255.255.0.
If you need to configure the sensor to another IP address (different than 192.168.170.4), you need to
use the $IPCN command (Cf. Appendix 1). This could be useful in case of multiple CHRocodile C
connection on a single computer.
Содержание CHRocodile C
Страница 1: ...1 CHRocodile C Compact sensor for non contact distance and thickness measurement Operation Manual...
Страница 3: ...3 Version Control Version Manual Date Type of Change 1 0 0 0 2016 09 20 Original edition...
Страница 13: ...13 Fig 2 2 Chromatic Confocal Imaging principle point sensor...
Страница 57: ...64 Appendix 2 Mechanical Plans 1 Optical probe mechanical plans b Probe 1mm a Probe 0 2mm...
Страница 58: ...65 Fig A2 1 Optical Head Mechanical plans a 0 2mm b 1mm c 4mm d 10mm c Probe 4mm d Probe 10mm...
Страница 59: ...66 2 CHRocodile C unit mechanical plans Fig A2 2 CHRocodile C sensor unit mechanical plans...