Trident Model PD765 Universal Input Meter Instruction Manual
7
SPECIFICATIONS
Except where noted all specifications apply to operation at +25°C.
General
DISPLAY
Trident: 0.56" (14 mm), Trident X2: 1.20" (31 mm),
Four digits (-1999 to 9999), automatic lead zero blanking.
DISPLAY INTENSITY
Eight intensity levels
DISPLAY
UPDATE RATE
Process/RTD: 3.7-5/second
Thermocouple: 1.8-2.5/second
OVERRANGE
Display flashes
9999
UNDERRANGE
Display flashes
-1999
PROGRAMMING
METHODS
Four front panel buttons, PC and MeterView
®
software, or
cloning using Copy function
NOISE FILTER
Programmable from 2 to 199 (0 will disable filter)
RECALIBRATION
All ranges are calibrated at the factory. Recalibration is
recommended at least every 12 months.
MAX/MIN
DISPLAY
Max/min readings reached by the process are stored until
reset by the user or until power to the meter is turned off.
PASSWORD
Programmable password restricts modification of settings.
NON-VOLATILE
MEMORY
All programmed settings are stored in non-volatile memory
for a minimum of ten years if power is lost.
POWER
OPTIONS
85-265 VAC, 50/60 Hz
90-265 VDC, 20 W max
or 12-36 VDC, 12-24 VAC, 6 W max
See table for power consumption
(*X: number depends on option)
Model Watts
PD765-6RX-0* 8
PD765-6RX-1, 2*
20
PD765-7RX-0* 6
FUSE
Required fuse: UL Recognized, 5 A max, slow blow
Up to 6 meters may share one 5 A fuse
ISOLATED
TRANSMITTER
POWER SUPPLY
One or two transmitter power supplies (Optional)
P or P1: 24 VDC
±
10% @ 200 mA max. (-1 option)
P1 & P2: 24 VDC
±
10% @ 200 mA & 40 mA max. (-2 option)
NORMAL MODE
REJECTION
64 dB at 50/60 Hz
ISOLATION
4 kV input/output-to-power line
500 V input-to-output or output-to-P1/P2 supplies
-6R5 & -6X5 models only: 100 V output-to-24 VDC supply
OVERVOLTAGE
CATEGORY
Installation Overvoltage Category II:
Local level with smaller transient overvoltages than Instal-
lation Overvoltage Category III.