Trident Model PD765 Universal Input Meter Instruction Manual
48
Program the Sensor Break Output Value (
SEbr
)
The sensor break value corresponds to the output signal generated
when the meter detects a sensor break for thermocouple and RTD in-
puts.
For example if there is an open thermocouple, the meter displays the
message “
open
” and the analog output goes to the programmed sensor
break value (e.g. 3.00 mA).
The sensor break value can be programmed from 0.00 to 23.99.
The typical output signal range is 1.00 to 23.00 mA (e.g. If sensor break
value is programmed to 0.00, the actual output will not be greater than
1.00 mA).
Analog Output when Display is Out of Range
The analog output reflects the display out of range conditions as follows:
Input Condition
Display
Analog Output
Underrange Flashing
-1999
3.00
mA
Overrange Flashing
9999
21.00
mA
Open TC or RTD
Flashing
open
Sensor break value
sEbr
Set Sensor
Break Value
scal
0
3
.00
Aout
Program
Menu
Press Enter to Accept Setting
Press Up to Set Digit Value
Press Right to Select Next Digit
Press Menu to Exit at any Time