Trident Model PD765 Universal Input Meter Instruction Manual
35
Calibrating the Meter (
Cal
)
To scale the meter without a signal
source, refer to Scaling the Meter
(
scal
), page 33.
The meter can be calibrated to display the process in engineering units
by applying the appropriate input signal and following the calibration
procedure.
The use of a calibrated signal source is strongly recommended to cali-
brate the meter.
Recalibrating Temperature Inputs (
Cal
)
Remember, the meter is
calibrated
at the factory prior to shipment. Re-
calibration is recommended at least every twelve months.
The
Calibration
(
CAL
) menu is used to
recalibrate
the thermocouple and
RTD inputs.
Allow at least 30 minutes warm-up time
before performing recalibration proce-
dure to ensure specified accuracy.
inp1
dis1
0
4.00
Set Display 1
Value
inp2
Calibrate
Input 2
Cal
Display
Flashes
Accepting
Input
See
Scaling the Meter
for Button
Functions Description