14 Density ("DENSITY")
40
14.7 Density of a porous solid "MODE SOLID POROUS"
This method is used to determine the density of a porous solid. The double beaker is required for this. The
thermometer is used to monitor the temperature of the test liquid in the container.
D
ensity determination using a density determination set
Order number see Accessories series 320
fig. 5
fig. 6
fig. 7
Density determination using below balance weighing
Order number see Accessories of the respectiv
series
fig. 5a
fig.6a
fig. 7a
Display
Key
Step
0,000
9
«
T
»
Taring
AIR
SET
Place the porous solid into the top beaker (fig. 5 resp 5a).
+
24,974
9
«
B
»
As soon as the reading is stable, it is saved and dis-
played for 2 seconds.
AIR
SET
Seal the pores on the solid with wax, for example. Now place the solid into the top beaker (fig. 6 resp 6a).
+
26,971
9
«
B
»
As soon as the reading is stable, it is saved and dis-
played for 2 seconds.
CLOSED PORES
SET
Place the solid into the bottom beaker (fig. 7 resp 7a)
+
16,984
9
«
B
»
As soon as the reading is stable, it is saved and dis-
played for 2 seconds.
LIQUID
SET
+
2,496
9.ccm
Solid density display
(provided it is within the 0,1 - 25.0 g/ccm range).
DENSITY
<->
+
2,05
%
«
C
»
Switch to the density index display
(if Index is ON).
INDEX
<->
0,000
9
«
T
»
Taring, the balance returns to the gram display and is
ready for the next determination.
AIR
SET