24
NT2-210018 Optical Spectrum
Analyzer User Guide
6176 E. Molloy Rd. East Syracuse, NY 13057 U.S.A. • [email protected] • 1-800-800-6652 • +1 315-431-7200 • www.ppc-online.com
A
BELDEN
BRAND
Optical Network
Solutions
12.2 Calibration and Verification
Periodic verification of the NT2-210018 is recommended to ensure your instrument remains within specification.
Although not imperative, we recommend a calibration and verification once a year to make certain the instrument
is functioning properly and performing to its rated specifications. Consult the factory for service.
12.3 NT2-210018 Adapter Replacement
The NT2-210018 is supplied with two easily interchangeable adapters per port, SC/FC. To change an adapter,
remove the two screws that hold the adapter in place, pull the adapter straight up from ferrule. It is suggested you
clean the exposed ferrule with appropriate cleanser and a lint-free wipe any time you replace the ferrule.
Note
In order to maintain a low loss fiber connection, care should be taken to adequately clean the ferrule of any
connector to be connected to the OSA. In the event that the port needs to be cleaned, the first step is to be
certain the instrument is off. We suggest the use of isopropyl alcohol and foam swabs specifically designed
for cleaning connectors accepting 2.5mm ferrules.
When replacing the adapter with one that does not have a chained protective cap, use the small screw in
place of the larger screw that retains the end of the chain to the adapter base.
Section 13 - Specifications
NT2-210018 Specifications
Wavelength Range
C-Band 1530 -1561nm (196.0 THz - 192.0THz)
Channel Spacing
50GHz, 100GHz
Wavelength Accuracy
±0.1nm
Channel Power Range
+10dBm to -50dBm
Absolute Accuracy
±1 dB
Max Composite Power
+28 dBm
PDL
±0.15dB
Optical Rejection Ratio
40dBc (@50GHz)
Measurement Time
< 1/2 Second
Readout Resolution
0.01dB
Return Loss
> 40dB
Optical Interface
Universal UPC (FC/SC)
Graphical Display
Bar graph and table view
Display
4 in. Touch color TFT
Dimensions
7.62” L x 3.88” W x 1.56” H (194mm L x 99mm W x 40mm H)
Weight
1.6 lbs
Battery
Rechargeable NiMH - 4 hours operating time
Power
100-240 universal US, GB, EU, AU Mains
Environmental
Operation -10°C to 50°C
Accessories Included
Universal power supply with mains for US, UK, CE and AU. Interchangeable FC and
SC adapters, CertSoft Software Suite,USB cable, manual on CD and rubber boot
Specifications are subject to change without notice