Check the sludge titration (dry weight).
2)
Clean the sensor head with a water jet, focussing on the insides of the optical measuring channel.
If you are unable to calibrate the SS sensor after these checks, seek advice from an after-sales service (see last
page).
6.3 Temperature sensor calibration error message information
6.3.1 0°C calibration error
If an error occurs during calibration in water at 0°C:
1)
Check the water temperature with a precision thermometer.
2)
Check the connection between the temperature sensor and ACTEON 2050.
If you are unable to calibrate the SS sensor after these checks, seek advice from an after-sales service (see last
page).
6.3.2 Ambient water calibration error
If an error occurs during calibration in water at ambient temperature:
1)
Check the water temperature with a precision thermometer.
2)
Check the connection between the temperature sensor and ACTEON 2050.
If you are unable to calibrate the SS sensor after these checks, seek advice from an after-sales service (see last
page).
CALIBRATION
ESC: Escape menu
OF TEMPERATURE SENSOR
CALIBRATION IMPOSSIBLE
Slope:
NON CORRECT
CALIBRATION
TEMPERATURE SENSOR
CALIBRATION IMPOSSIBLE
Offset:
NON CORRECT
ESC: Escape menu