User manual
PONSEL
Référence : NOTICE_ACTEON_2050_v001
V : 002
12/67
It is best to use the elbowed sensor-holder perch with its own installation system when submerging the sensor in
a tank.
Figure 5 – Elbowed Sensor-Holder Perch
Comment:
Use elbowed sensor-holder perches in heavily soiled tanks to prevent fibre build-up on the perch.
If there is a directional flow, the sensor's optical slit should be pointing in the direction of the flow.
Installing the SS sensor in the nozzle:
The standard sensor comes sealed to a 10-metre cable connection, making it submersible up to a several bars of
pressure.
1)
Completely unscrew the tightening seal located on the sleeve of the SS sensor beforehand.
2)
Pass the sensor cable through the bottom of the perch (widest end) until it comes out of the other end.
3)
Slide the cable through the perch until the SS sensor sleeve fits into the bottom of the perch.
4)
Slot the SS sensor sleeve insert in the lateral hole located in the lower end of the perch. Tighten the
lock-nut until the sensor is secured in the perch (do not over tighten).
5)
Set the perch on the edge of the tank or on the bridge.
SS sensor
Lock-nut
O-Ring seal
Structure for holding sensor
Elbowed Sensor-Holder Perch
SS sensor
30cm
minimum
Liquid flow direction
Liquid level